PSR 11-55 Replacement oxygen sensor for Hudson 5577, 6477, and 5590

OEM Equipment:
Hudson 5577 Analyzer
Hudson 6477 Analyzer
Hudson 5590 Monitor

PSR 11-55

Description

Measuring Range 0-100%
Accuracy 1 +/-2% of Full Scale
Signal Output 2 7-14 mV
Linearity +/-2% of Full Scale
Response T90 9 sec
Temperature Coefficient Compensated
Operating Temp 0 to 45°C
Recommended Storage 3 0 to 25°C
Temperature Coefficient Compensated
Shelf Life 4 6 months
Humidity Non-condensing 0-99% RH
Expected Life 2 42 months
Warranty 5 14 months
Electrical Connector 4 Pin Molex

* Conditions – Specification validated during design and in pursuit of improvement are subject to change without notice
Note:

  • 1) At constant temperature and pressure.
  • 2) In air (20.9% oxygen) at 25°C and 1 atm. Expected life varies inversely with changes in O2 concentration, temperature, pressure.
  • 3) Sensor may be stored up to 55°C on an intermittent basis, for example, during transportation.
  • 4) In original Package at 25°C and 1 atm.
  • 5) Under normal operating conditions, the sensor is warranted to be free of defects in material and workmanship for the specified period provided the sensor is properly installed and operated. The sole remedy for sensor determined to be defective by Analytical Industries Inc. is limited to replacing the sensor. Analytical Industries Inc. will not be liable for buyer???s negligence, misapplication, abuse or accident.

Reviews

There are no reviews yet.

Be the first to review “PSR 11-55 Replacement oxygen sensor for Hudson 5577, 6477, and 5590”

Your email address will not be published. Required fields are marked *

Categories
Contact

Get single parts

Get tech support