PSR 11-33 Replacement galvanic oxygen sensor for use in Hamilton Amadeus and Hamilton Veolar.

OEM Equipment:
Hamilton Amadeus
Hamilton Ventilator

PSR 11-33

Description

Measuring Range 0-100%
Accuracy 1 +/-2% of Full Scale
Signal Output 2 165-235 uA
Linearity +/-2% of Full Scale
Response T90 30 sec
Temp Coefficient Compensated
Operating Temp 0 to 45°C
Recommended Storage 3 0 to 25°C
Temperature Coefficient Compensated
Shelf Life 4 6 months
Humidity Non-condensing 0-99% RH
Expected Life 2 24 months
Warranty 5 12 months
Electrical Conn 2 ring PCB

* Conditions – Specification validated during design and in pursuit of improvement are subject to change without notice
Note:

  • 1) At constant temperature and pressure.
  • 2) In air (20.9% oxygen) at 25°C and 1 atm. Expected life varies inversely with changes in O2 concentration, temperature, pressure.
  • 3) Sensor may be stored up to 55°C on an intermittent basis, for example, during transportation.
  • 4) In original Package at 25°C and 1 atm.
  • 5) Under normal operating conditions, the sensor is warranted to be free of defects in material and workmanship for the specified period provided the sensor is properly installed and operated. The sole remedy for sensor determined to be defective by Analytical Industries Inc. is limited to replacing the sensor. Analytical Industries Inc. will not be liable for buyer???s negligence, misapplication, abuse or accident.